Model 1175:    Isolated/Specialized Thermocouple Interface

Datasheet    XIF File

Description

The Model 1175 is an Isolated Thermocouple Network Interface with unique and specialized features.

The Model 1175 has been tested with XCO Corporation’s FTLD and CT2 Continuous Thermocouple sensors.  These are unique sensors that could extend for several hundred feet and measure the highest temperature along their entire length.  The Model 1175 is particularly well suited for use with these sensors due to its excellent performance, very high input impedance, and extremely low input current.   Please visit the www.xco.com website for more information on this product. Finally, the Model 1175’s Thermocouple input is isolated from the DC input power supply.  This allows it to be used with grounded thermocouple as well as non-grounded types.

Key Features

bullet

Very compact design, available in DIN Rail and NEMA 4X enclosures.

bullet

FTT-10A network interface with temperature information available using multiple Standard Network Variable, SNVT, formats.

bullet

Wide operating voltage range of 10 to 32 VDC with a typical input current of 55 mA. 

bullet

High quality removable terminal blocks on all power and network connections.

bullet

Cold junction compensation via Class A platinum RTD.

bullet

Extremely low input current, 10 Pico Amps typical, and very high input impedance of 20 M Ohm

bullet

Electronics have extremely high temperature stability and low drift.

bullet

Analog to Digital converter has 24-bit resolution.

bullet

Two status LEDs provide quick visual indication of runtime status conditions.

bullet

Compatible with XCO’s FTLD and CT2 Continuous Thermocouple sensors.

bullet

-40 to 85 C operating temperature range with maximum input voltage of 24 VDC.

bullet

Initial support for K-type, -200 to 1372 C, and planned support of a wide variety of standard thermocouple types.

bullet

High precision thermocouple polynomial

 